S.C.R. WAGON WORK SHOP, GUNTUPALLI

COMPUTERIZED ELECTRONIC DOUBLE CUP MEASURING SYSTEM (LVDT)

Procedures for Computerized Electronic Gauging System

SETTING UP THE GAUGE SYSTEM:

     1. Open the PC cover. Fit the interface card 1711L in the PCI slot. Confirm the 230 ± 10 %V, 50 ± 10%Hz
        power supply. If power fluctuation is there then it is re-commanded that the standard voltage
        stabilizer be incorporated.
     2. Connect the PC to the mains power supply. Also confirm that the probe box is connected to the mains.
     3. Connect the probe box to the interface card which is fitted in PC using 25 — 37 pin connector.
     4. Insert probes to the probe holder.
     5. Connect the respective plugs of probe to the probe box.

SOFTWARE INSTALLATION:

     1. Insert the software CD into the CD drive.
     2. Double click on CD drive details.
     3. First install " mdac & le5"
     4. Double click on the set-up.
     5. Follow the subsequent instructions i.e. next .
     6. Restart the PC.
     7. Install the drivers of Advantech card.
     8. Insert the CD named Advantech into the CD-DRIVE.
     9. Install the Device manager.
    10. Install the PCI drivers.
    11. Restart the PC.
    12. Go to start of the Window.
    13. Select programs.
    14. Go to Advantech Device drives.
    15. Select Advantech Device manager.
    16. Choose "PCI-1711L".
    17. Click on "Test".

PROBE LOADING PROCEDURE:

     1. Insert probes into the probe holder.
     2. Connect the respective plugs of probe to the probe box as per the marking made.
     3. Insert the Master in the fixture.
     4. Now physically load the probes on the master.
     5. Set all the probes in the probe-loading window (in Software) near about zero reading.
     6. Tight the probes by Allen screw (Do not over tight the probes)